The RI8611A Cassini V93K CTH Infrastructure system can operate in a wide range of environmental conditions. Depending on the system configuration, some specifications may not apply or have individual requirements. No special plumbing or ventilation is required. Customer must provide items designated with "NOT PROVIDED" label. All values are subject to change without notice.
- Height: 0.256 m (10.08 in.)
- Depth: 1.056 m (41.61 in.)
- Width: 0.754 m (29.72 in.) Includes TIMs
- System Weight: 79 Kg ( 174 lbs.) (8 TIMs)
- Max Shipping Weight: 105 Kg (230 lbs.)
- Electrical via 48V AC-DC Power Adapter:
- IEC320-C13(f) [PSU side is IEC-320 C14(m)]
- Input AC Power Requirements: 3.8A/115VAC or 2A/230VAC
- Max Inrush Current: 95A/115VAC or 120A / 230VAC
All Electrical Input:
- Input voltage and frequency ranges supported automatically.
- 220 V or 110 V
- 50 Hz or 60 Hz AC, Single Phase Sine-wave input.
Fixture: (when docking with a Handler or Prober)
- Always consider the Fixture height when determining minimum clearance of the Test Head.
- Height: 195.6 mm (7.7 in.) - includes clearance for alignment pins (varies based on Model/Design)
- Depth: 266.7 mm (10.5 in.)
- Width - 16 slot test head: 558.8 mm (22 in.)
- Force air option may be included, documentation provided separately.
Environment:
- Operating Air Temperature: 10° to 35°C (50° to 95°F)
For best calibration performance, a consistent ambient room temperature of ± 5°C is recommended. - Humidity: 8% to 80%
- Docking options based on Handler/Prober Type: Hard dock (RIK0070A, HNDPLATE), Soft dock (Cables)
- Handler "Pod" per handler/prober per system: GPIB (RIK0263A), Serial (RIK0266A), Parallel (RIK0267A) - N/A for RI8611A
Requires factory RI configuration and programming, send detailed handler control logic to "[email protected]" prior to ship date. - Handler Pod Interface Cable: Parallel 25 Pin Female D-Sub (Custom MFSHLX1A), Serial Female RS232, or GBIP cable. NOT PROVIDED
- Workspace surface not needed for Cassini V93K CTH (RI8611A)
- Workspace should be located within reach of the USB control cable that is 3.3m or (10 ft, 10 in.) long from the V93K Testhead.
- Workspace (cart) may needed to hold monitor, keyboard and mouse. NOT PROVIDED