Due to supply constraints, the RI8567B4 TIM has been revised to use Apria switches instead of CMC switches. These Apria switches have a different 50 Ω match characteristic.
- Applies to RfMeasure TIMs shipped after May 2024.
 - See the Table of RI8567B TIMs for affected serial numbers.
 - This document provides best‑practice configuration steps to maintain consistent match performance across hardware revisions.
 
Symptom
- Mismatch ripple appears in calibrated Fixture paths, where power seems to add or subtract at different frequencies.
 - This effect is repeatable and predictable, so final measurements remain valid. Calibration compensates for both source and receiver paths.
 - Fixture calibration is hardware‑generation dependent:
 - B3 and earlier: smoother calibration curves
 - B4 and later: ripple visible (see Figure 1) unless best-match configuration is used
 - Following the best‑match procedure aligns calibration data across TIM hardware generations.
 
Recommendations
- Share this document with engineers comparing Fixture calibration data between B3‑and‑earlier and B4‑and‑later TIMs.
 - Avoid swapping TIMs between hardware generations. If swapping is required, perform a full Fixture calibration.
 - No software changes are needed, but document and communicate the expected calibration differences (see Figure 1).
 - If source match issues are identified, follow the procedure “Update Testplans for Best Match S‑Parameter Measurements”.
 - Always update both Fixture Calibration Test Plans and Device Test Plans to use the same switched paths.
 - Identify TIM hardware version by:
 - Checking in Cassini software, or
 - Comparing the serial number sticker with the Table of RI8567B TIMs.
 
Procedure – Update Testplans for Best Match S‑Parameter Measurements
- In Cassini, open the Fixture Calibration Testplan (Test > Cal Plans… or System > Test Objects).
 - Choose Options > Display States.
 - Locate RfMeasure1 Receive Mode entries with value = sPar.
 - This routes signals through switch M26 to open port on switch M29, which causes the poor match. (See Figure 5)
 - For RfMeasure1 Rf2 entries with value = Rec:
 - Double‑click to view the test panel.
 - Change RfMeasure | Rf2 from Receiver → Noise (Noise Source + 6 dB pad) or if RfMeasure | Rf3 is set to Receiver, change to Load.
 - Repeat for all sections in the States window.
 - Compile and run the Testplan. Verify reduced mismatch ripple in Fixture calibration data.
 - Save changes (File > Save Guru).
 - Apply the same procedure to Device Testplans.
 
Identifying TIM Hardware Revision
Using Equipment Pool:
- Launch latest Dev or Production Cassini app.
 - Go to System > Equip > Nodes.
 - Select RfMeasure1.
 - Check BaseInst => attribute. The last two digits = hardware revision (e.g., “B7”).
 
- Launch latest Dev or Production Cassini.
 - Go to System > Tester.
 - Choose Tester > Create Config Log….
 - Save to D:\Shared or a network folder.
 - In the log, find baseInst= under “Fixture Connections” for RfMeasure1.
 
Detailed Explanation of Affected Tests
- RfMeasure | Rf2 button
 - Default: set to Receiver, controlling switch M29 → M26.
 - Fix: change to Noise (M29 → Noise Source + 6 dB pad).
 - RfMeasure | Receive Mode button
 - Default: set to s‑parameters, controlling switch M2 → M26.
 - Routes RF2 (C7) input to an open, disconnected path.
 - Example: When Output Port = RF2, measurements follow a parallel path through coupler + switches → Receiver.
 
Figure 1: Example Fixture Cal Data showing Mismatch Ripple (>=B4 Reflective on left <=B3 Absorptive on right)

Figure 2: Before and After Best Match Change
  
Figure 3: Apria "Rectangle" Switches in rev B4 and after.

Figure 4: CMC "Octagon" Switches in rev B3 and earlier.

Figure 5: Block Diagram Showing Testplan Change (Signal Terminated at Noise Src vs Open)

Table of RI8567B TIMs
Affected instruments are HW Version "B4" and later.
Switch  | HW Ver  | Serial Numbers | 1st Ship | 
CMC  | B1  | RIF3R14A, RLPS9C0A | 2011-05 | 
CMC  | B2  | RMKUYU0A, RMLNAU0A, RMV5LE0A | 2012-04 | 
CMC  | B3  | PR0NJH0A, RP8MX50A, RP955T0A, RPNXLZ0A, RPRRFE0A, RPSK860A | 2013-05 | 
Apria  | B4  | RPVSJT0A, RPX5YY0A | 2014-05 | 
Apria  | B5  | RIR1F6W0, RR0R0S0A, RR23SR0A, RR2Y360A | 2014-07 | 
Apria  | B6  | RR4X650A, RR5E510A | 2014-08 | 
Apria  | B7  | RR7WBG0A, RR9BVY0A, RRFMBW0A, RRLS4R0A, RTG2CJ0A, R2G2E10A | 2017-01 | 
Apria  | B8  | RVR7YN0A (MISLABELED MVSW081A), RIUGZEK0, RUWA3V0A, RUWA440A, RVCFY80A, RVCG230A, RVR80g0A, RVUM7S0A, RVUMHX0A, RW392B0A | 2018-06 | 
Apria  | B9  | RWFE980A | 2020-06 | 
Apria  | BA  | RWNUHK0A, RXCCYL0A | 2021-03 | 
As of 2025-November, rev BA is the latest revision.
TIM Block Diagrams
RI8567BA_Test_Set_12GHz w Cable Lengths.pdf<-- Reflective Switches by Apria